Figure 2: Different shapes of pulse wave signal depending 
on the sensor pad position: ● — pad is located accurately 
above the artery axis, ■. ♦ — pad is moved left and right 
from the center of the radial artery, respectively. 
This leads to two important conclusions  relating 
to positioning the sensor. First, the central measuring 
chamber must be located accurately above the artery 
axis projection. Second, the artery must be pressed by 
the sensor to underlying tissues in a way its axis does 
not  waver  upon  pulsation.  These  observations 
stimulated  the  authors  to  practically  realize  the 
"targeting"  method  similar  to  the  lateral  signals 
equalizing  approach  used  in  radar  technique. 
According  to  that,  the  authors  designed  a  blood 
pressure  monitoring  sensor  whose  main  part  is  the 
measuring block with three separate chambers. Each 
chamber has its own independent measuring output. 
During  the  operation,  the  channel  nozzles  are 
positioned  in  a  row  transversely  against  the  artery. 
With that, the dimensions of the area of contact must 
be  designed  to  ensure  that  all the  three  nozzles  are 
located above the subject artery during the operation. 
Figure 3 gives the sketch of the sensor and the result 
of concurrent three-channel pulse wave measurement 
(the sensor located above the radial artery).  
 
 
Figure  3:  Three-chamber  sensor  enabling  local-
compensation  blood  pressure  measurement  (A);  and 
synchronous three-channel pulse wave chart (B). 
The comprehensive specification of the sensor is 
provided  in  the  patent  (Mansurov  et  al.,  2018).  In 
terms of  the proposed design, the main task of  side 
channels is to ensure proper positioning of the central 
measurement  pad.  With  the  correct  position  of  the 
measuring  unit,  the  calibrated  signals  on  the  side 
channels  (Figure  3)  coincide  or  slightly  differ  from 
each other. It can be neglected that artery walls under 
side  channels  cannot  be  fully  unloaded,  so  the 
pulsation response in those channels is significantly 
distorted. It is only important that upon the equality 
of  those  signals,  the  central  chamber  is  positioned 
accurately  above  the  artery  axis  ("targeted")  —  in 
such a position, that its signal will be a non-distorted 
copy of arterial pressure (Antsiperov et al., 2018). 
The methodology of measuring blood pressure by 
the three-chamber pneumatic sensor is tightly related 
to the described design features. At the first stage, just 
before the measurement, the location of the artery is 
determined by palpation. Then the sensor is applied 
onto  that  place  so  that  measuring  chambers  are 
positioned  in  a  row  transversely  against  the  artery 
(Figures 2 and 3). Then, manually moving the sensor 
along this direction (transversely against the artery), 
the physician should find a position in which signals 
of side channels are as equal as possible. After that, 
the  measurement  unit is  pressed  against the  skin  so 
that the contact area under the central pad became flat, 
but  without  the  artery  occlusion  (applanation 
principle). For the radial artery case, the criterion of 
the  best  position  was  experimentally  determined. 
According  to  it,  the  signal  amplitude  of  the  central 
channel must be about twice as high as the equalized 
amplitudes of the side channels.  
The  problem  of  positioning  the  sensor  on  the 
radial  artery  was  addressed  via  designing  a 
monolithic  three-chamber  sensor  that  reflected  the 
problem specifics. Three chambers are made in a rigid 
flat  surface  (1.8  mm  increment)  along  the  line 
perpendicular  to  the  artery  axis  projection  on  the 
sensor plane. The chambers are independently fed by 
the  air  from  the  receiver  through  individual  air 
throttles. This way the pressure on the sensor surface 
can  be  measured  concurrently  and  independently  at 
three  points  (0.8  mm  in  diameter).  To  estimate  the 
thickness  of  the  air  cushion  underneath  the  plane 
surface,  the  current  air  consumption  rate  was 
measured  and  amounted  to  ~  0.5  cm
3
/s  (with 
accounting possible leaks). By the average pressure 
difference of 100 mm Hg, the discharge flow speed in 
the  hole  is  about  140  m/s.  The  flow  cross-section 
thereby totaled 500/140,000 = 0.0036 mm
2
 or 0.0012 
mm
2
 for each camera. Assuming that air is discharged 
within the half of the chamber hole perimeter (cleft 
length is ~ 1 mm), the cleft width should be ~ 1 µm. 
The sensor of this type can be used for measuring 
parameters  of  radial  and  other  arteries  (carotid, 
temporal,  etc.).  However,  such  a  sensor  positioning 
algorithm  enabling  adequate  quantitative  blood 
pressure measurement is  developed and  empirically 
confirmed  only  for  positioning  the  sensor  on  the 
radial artery.