
CT scanner X-ray beam leads to substantial 
distortions in dose estimates despite the fact that the 
last method takes into account beam rotation
.TLDs 
(LiF:Mg, Ti; 4.5±0.12 mm in diameter, 0.9±0.1 mm 
thickness) from chosen group were placed inside a 
container in which 22 – 24 sensors are housed 
simultaneously. As a reference dosimeter Unfors Xi 
with pencil type ionization chamber has been used. 
Container with TLDs as well as Unfors Xi detector 
were installed at the central and periphery holes of 
CTDI phantom located in the centre of scanner 
gantry and after this several exposures for one tube 
rotation and with fixed table position have been 
made. Multiple exposures are needed in order to 
reliably exceed the sensitive threshold of 
thermoluminescent dosimeters. Considering 
reference device (Unfors Xi) readings and TLDs 
measurements (estimates of an integral under 
absorbed dose distribution curve on ±50 mm 
interval) two values of the weighted CTDI have 
been calculated and their ratio gives the calibration 
factor for testing TLDs group.  
Effective doses on DLP basis have been 
calculated in respect that conversion factor for adult 
chest studies in accordance with appropriate 
document is equal to 0.017 mSv·mGy
-1
·cm
-1 
(European Commission’s Study Group, 1999).  
An adult (hermaphrodite) anthropomorphic body 
phantom consists of 25 mm elements simulating 
human body organ sections with different densities. 
This phantom with TLDs (about 100 sensors) placed 
inside was chosen for reference measurements. One 
phantom section with thermoluminescent sensors 
housed in is shown on Fig.1, as an example. TLDs 
distribution in phantom was the following: red bone 
marrow – 7;  colon – 5; lungs – 24; stomach – 4; 
breast – 2; gonads – 2; bladder – 2; liver – 5; 
esophagus – 8; thyroid – 6; skin – 11; bone surface – 
10; remind organs – 13. Using an anthropomorphic 
phantom+TLDs, effective dose evaluation has been 
made on the basis of weighting factors for different 
organs and tissue presented in International 
Commission on Radiological Protection (ICRP) 
documents: ICRP Publication 60 and ICRP 
Publication 103 recommendations (ICRP, 1991); 
(ICRP, 2007). 
3 RESULTS AND DISCUSSION 
The right parts of absorbed dose distribution curves 
received at the central and periphery holes of 
doubled CTDI “body” phantom are presented on 
Fig.2 and Fig.3 respectively. Fig.4 shows the right
parts of weighted distribution curves. Full length 
weighted curves have been used for weighted CTDI 
calculation on the basis of integrals under the curves 
on the interval from –200  to +200 mm). Weighting 
has been made using weighted factors 1/3 and 2/3 
for centre and periphery zones respectively. It 
follows from the received data that the integrals 
under the absorbed dose distribution curves at the 
central hole of doubled CTDI phantom (in the range 
of ±200 mm) differ from the integrals on the interval 
from –50 to +50 mm (traditional CTDI method) 
approximately by (53 – 65)% for both scanners. For 
periphery holes this difference is not so significant 
and is about (11 – 13)%. Mentioned above means 
that ignoring of scattered radiation outside ±50 mm 
zone along z-axis leads to underestimation of the 
effective dose up to approximately 20% in average 
when using the traditional method on the basis of 
single CTDI "body" phantom and 100 mm 
ionization chamber (weighted CTDI) in case of 
collimation 32 – 40 mm. 
 
 
Figure 1: Anthropomorphic phantom section with TLDs  
housed in. 
 
Figure 2: The right part of absorbed dose distribution at 
the centre hole of doubled CTDI phantom.
 
BIODEVICES2013-InternationalConferenceonBiomedicalElectronicsandDevices
92