Authors: Eugenio Mattei ; Giovanni Calcagnini ; Michele Triventi ; Federica censi and Pietro Bartolini

Affiliation: Italian National Institute of Health, Italy

Keyword(s): Magnetic Resonance Imaging, Pacemaker, Specific Absorption Rate, Human Visible Dataset.

Related Ontology Subjects/Areas/Topics: Bio-Electromagnetism ; Biomedical Engineering ; Biomedical Instrumentation ; Biomedical Instruments and Devices ; Implantable Electronics

Abstract: Numerical simulations were performed to evaluate the Specific Absorption Rate (SAR) induced at the tip of a pacemaker (PM) implant by the 64 MHz radiofrequency (RF) field used in 1.5T Magnetic Resonance Imaging (MRI) procedures. The analysis was performed by using a commercial FDTD software (SEMCAD X, SPEAG, Switzerland) and aimed at the evaluation of the impact that the patient ‘s morphology has on the induced local SAR at the implant tip. In particular three human phantoms were studied: a 34-year old man model, a 26-year old woman, and a 6-year old boy. The three phantoms reproduce more than 70 tissues of the human body with a resolution of 1 mm. Inside each phantoms, realistic implant configurations were modelled, considering both left and right pectoral implants, and atrial and ventricular stimulations. The local SAR values at the lead tip was compared for the three phantoms and sensible differences were observed: with a RF excitation set to produce an whole-body average SAR of 2 W/kg without any implants, local SAR values ranged from 641W/kg (woman model – right ventricular implant) to 3 W/kg (boy model – left atrium implant). We also observed that, in general, ventricular implants showed a higher SAR compared to atrial ones, as well as right pectoral implants compared to left ones. However, not always a higher implant area or a longer lead path implied higher SAR at the tip, indicating the coupling mechanisms between the implant and the RF field are likely to be more complex that the only area-dependent induction law. (More)

CC BY-NC-ND 4.0

Sign In Guest: Register as new SciTePress user now for free.

Sign In SciTePress user: please login.

PDF ImageMy Papers

You are not signed in, therefore limits apply to your IP address 18.220.59.69

In the current month:
Recent papers: 100 available of 100 total
2+ years older papers: 200 available of 200 total

Paper citation in several formats:
Mattei, E.; Calcagnini, G.; Triventi, M.; censi, F. and Bartolini, P. (2011). MRI-INDUCED SAR ON PACEMAKER LEADS - Numerical Simulations on Three Human Phantoms. In Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2011) - BIODEVICES; ISBN 978-989-8425-37-9; ISSN 2184-4305, SciTePress, pages 135-139. DOI: 10.5220/0003131501350139

@conference{biodevices11,
author={Eugenio Mattei. and Giovanni Calcagnini. and Michele Triventi. and Federica censi. and Pietro Bartolini.},
title={MRI-INDUCED SAR ON PACEMAKER LEADS - Numerical Simulations on Three Human Phantoms},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2011) - BIODEVICES},
year={2011},
pages={135-139},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0003131501350139},
isbn={978-989-8425-37-9},
issn={2184-4305},
}

TY - CONF

JO - Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2011) - BIODEVICES
TI - MRI-INDUCED SAR ON PACEMAKER LEADS - Numerical Simulations on Three Human Phantoms
SN - 978-989-8425-37-9
IS - 2184-4305
AU - Mattei, E.
AU - Calcagnini, G.
AU - Triventi, M.
AU - censi, F.
AU - Bartolini, P.
PY - 2011
SP - 135
EP - 139
DO - 10.5220/0003131501350139
PB - SciTePress