loading
Papers Papers/2022 Papers Papers/2022

Research.Publish.Connect.

Paper

Paper Unlock

Authors: Rajeev Ranjan ; Bibin John ; Dipal Gosh ; Soumil Kumar ; Lait Abu Saleh ; Dietmar Schroeder and Wolfgang H. Krautschneider

Affiliation: Hamburg University of Technology, Germany

Keyword(s): Aneurysm Implant, Blood Pressure Measurement, Stent Graft, Forward Error Correction.

Related Ontology Subjects/Areas/Topics: Biomedical Engineering ; Biomedical Instruments and Devices ; Devices ; Health Monitoring Devices ; Human-Computer Interaction ; Implantable Electronics ; Low-Power Design ; Physiological Computing Systems

Abstract: Aneurysm implant requires multiple pressure sensors integrated into the stent graft for recording pressure profile inside an aneurysm sac. This paper details the design of an ASIC which receives power wirelessly using inductive coupling, fetches pressure data from a seperate chip and transmits data wirelessly. Besides, the ASIC extends the functionality of the external pressure measurement chip to support up to four pressure sensors by using an analog multiplexer. Error detection and forward error correction are implemented to improve data integrity for wireless data transmission. The design has been implemented and tested in 350 nm CMOS technology.

CC BY-NC-ND 4.0

Sign In Guest: Register as new SciTePress user now for free.

Sign In SciTePress user: please login.

PDF ImageMy Papers

You are not signed in, therefore limits apply to your IP address 3.149.251.155

In the current month:
Recent papers: 100 available of 100 total
2+ years older papers: 200 available of 200 total

Paper citation in several formats:
Ranjan, R.; John, B.; Gosh, D.; Kumar, S.; Saleh, L.; Schroeder, D. and Krautschneider, W. (2016). Wireless Energy and Data Transmission ASIC for Blood Pressure Measurement in an Aneurysm Implant. In Proceedings of the 9th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2016) - BIODEVICES; ISBN 978-989-758-170-0; ISSN 2184-4305, SciTePress, pages 256-262. DOI: 10.5220/0005815002560262

@conference{biodevices16,
author={Rajeev Ranjan. and Bibin John. and Dipal Gosh. and Soumil Kumar. and Lait Abu Saleh. and Dietmar Schroeder. and Wolfgang H. Krautschneider.},
title={Wireless Energy and Data Transmission ASIC for Blood Pressure Measurement in an Aneurysm Implant},
booktitle={Proceedings of the 9th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2016) - BIODEVICES},
year={2016},
pages={256-262},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0005815002560262},
isbn={978-989-758-170-0},
issn={2184-4305},
}

TY - CONF

JO - Proceedings of the 9th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2016) - BIODEVICES
TI - Wireless Energy and Data Transmission ASIC for Blood Pressure Measurement in an Aneurysm Implant
SN - 978-989-758-170-0
IS - 2184-4305
AU - Ranjan, R.
AU - John, B.
AU - Gosh, D.
AU - Kumar, S.
AU - Saleh, L.
AU - Schroeder, D.
AU - Krautschneider, W.
PY - 2016
SP - 256
EP - 262
DO - 10.5220/0005815002560262
PB - SciTePress