Design and Implementation of Monitor Tester for Validation of ECG Signals

Vladimir Kasik, Marek Penhaker, Jan Kubicek, Martin Cerny, Martin Augustynek, David Oczka, Alice Krestanova

2019

Abstract

The paper deals with a design and consequent realization of a monitor tester for validation of the ECG devices, and patient’s monitors for the ECG measurement. In the paper, we describe the electrical design of the plug-in module, which is intended for a transmission of the digital data from the computer. Data are sent by using the controlling SW connected to the parallel port in the plug-in module. An essential part of the proposed module is the three-channel 12 bit D/A converter AD 7398. This converter is able to simultaneously send data from all the three outputs. The converter output is unipolar in the volts units. For this reason, the transmitted signals are adjusted by using the differential amplifiers in order to convert them to the bipolar. The signal’s amplitudes are decreased to the millivolts level by using the voltage dividers. On the base of this solution, the Einthoven ECG leads may be detected.

Download


Paper Citation


in Harvard Style

Kasik V., Penhaker M., Kubicek J., Cerny M., Augustynek M., Oczka D. and Krestanova A. (2019). Design and Implementation of Monitor Tester for Validation of ECG Signals. In Proceedings of the 12th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2019) - Volume 1: BIODEVICES; ISBN 978-989-758-353-7, SciTePress, pages 127-133. DOI: 10.5220/0007248001270133


in Bibtex Style

@conference{biodevices19,
author={Vladimir Kasik and Marek Penhaker and Jan Kubicek and Martin Cerny and Martin Augustynek and David Oczka and Alice Krestanova},
title={Design and Implementation of Monitor Tester for Validation of ECG Signals},
booktitle={Proceedings of the 12th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2019) - Volume 1: BIODEVICES},
year={2019},
pages={127-133},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0007248001270133},
isbn={978-989-758-353-7},
}


in EndNote Style

TY - CONF

JO - Proceedings of the 12th International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2019) - Volume 1: BIODEVICES
TI - Design and Implementation of Monitor Tester for Validation of ECG Signals
SN - 978-989-758-353-7
AU - Kasik V.
AU - Penhaker M.
AU - Kubicek J.
AU - Cerny M.
AU - Augustynek M.
AU - Oczka D.
AU - Krestanova A.
PY - 2019
SP - 127
EP - 133
DO - 10.5220/0007248001270133
PB - SciTePress