LOW-INVASIVE HEATING AND TEMPERATURE
MEASUREMENT METHOD FOR HYPERTHERMIA TREATMENT
USING THE METAL COATED FERROMAGNETIC IMPLANT
WITH LOW CURIE TEMPERATURE
Kazutaka Mitobe and Noboru Yoshimura
Faculty of Engineering and Resource Science, Akita University, 1-1 Tegatagakuencho, Akita, Japan
Keywords: Hyperthermia, Cancer, Low-invasive, Induction Heating, Wireless Temperature Measurement System.
Abstract: Hyperthermia has been used for many years to treat various types of malignant tumor because tumor cells
are more sensitive to temperature in the range of 42-45°C than are normal tissue cells. In this study, we aim
to develop the local heating method and monitoring method for hyperthermia using the Ferromagnetic
Implant with Low Curie Temperature (FILCT) under high frequency magnetic field. The heat generation
inefficiency of FILCT causes a barrier in order to use this method as a bedside tool. In this research work,
we coated the FILCT with a metal material in order to improve the heat generation efficiency. The magnetic
permeability of FILCT decreased around the Curie temperature; therefore we can use FILCT as a thermal
probe by measuring of the changing vector of the magnetic flux at the outside of human body
noninvasively. In this paper, we describe about the experimental setup and in vitro experimental results.
1 INTRODUCTION
Hyperthermia has been used for many years to treat
various types of malignant tumor because tumor
cells are more sensitive to temperature in the range
of 42-45°C than are normal tissue cells (Cavaliere,
R., 1967 and Overgaard, K., 1972). There are a lot
of heating methods for hyperthermia. The most
commonly used method of heating in clinical
settings is capacitive heating using a radiofrequency
electrical field (Ishida, T., 1980, Abe, M., 1986 and
Hiraoka, M., 1994). The great advantage of
capacitive heating is that it is non-invasive.
However, this method can cause excessive heating
of the fat layer and is not suitable for site-specific
hyperthermia because it is difficult to selectively
heat only the local tumor region to the intended
temperature without also damaging normal tissue.
Moreover, this method needs an invasive
thermometer to measure the temperature of tumor
region. To overcome the disadvantages of capacitive
heating for the treatment of tumors, attempts have
been made to use inductive heating with magnetic
implants (Jordan, A., 1993 and Shinkai, M., 1996).
In 1982, Matsuki et al. developed a soft-heating
method using the ferrite implant with low Curie
point (Masaki, H., 1982). The merit of the soft-
heating method is that it needn’t measure the
temperature as for the heat generation is
automatically stopped up to Curie point of the
implant. We previously described a system in which
hyperthermia was induced using ‘ferromagnetic
implant with low Curie temperature (FILCT)’ low
enough to mediate automatic temperature control,
and demonstrated its antitumor effect in a mouse
melanoma model (Saito, H., 2008 and Ito, A. 2009).
However, the improvement of the heating efficiency
was necessary because much energy had been
required to heat the target tumor.
In order to solve this problem, we have mixed
the high conductivity metal material with FILCT.
At the same time, we have used FILCT as a probe of
thermometer. Thus a less invasive heating and
temperature measurement system was developed
based on this idea. A key advantage of this system is
that it can distinguish whether the target temperature
has reached the Curie temperature or not by non
contact method. On the other hand, the infrared (IR)
thermography, which is the most popular noncontact
thermometer, can measure the skin temperature but
not the internal body temperature. We have
developed the experimental setup for the less
341
Mitobe K. and Yoshimura N..
LOW-INVASIVE HEATING AND TEMPERATURE MEASUREMENT METHOD FOR HYPERTHERMIA TREATMENT USING THE METAL COATED
FERROMAGNETIC IMPLANT WITH LOW CURIE TEMPERATURE.
DOI: 10.5220/0003152803410344
In Proceedings of the International Conference on Biomedical Electronics and Devices (BIODEVICES-2011), pages 341-344
ISBN: 978-989-8425-37-9
Copyright
c
2011 SCITEPRESS (Science and Technology Publications, Lda.)
invasive thermometry (Mitobe, K., 2008). In this
paper, we have assessed the appropriateness of this
system by the experiment of temperature
dependence and heating efficiency using the ‘Au
coated FILCT’.
2 EXPERIMENTAL SETUP
Ferromagnetic material with high magnetic
permeability attracts magnetic flux. If the
temperature of ferromagnetic material increases over
Curie temperature, then the magnetic permeability
decreases immediately. As a result, the material can
not attract magnetic flux. Figure 1 shows
fundamental concept of non-invasive thermometry.
The temperature of FILCT can be measured
noninvasively by the shift of vector of magnetic flux.
The Curie temperature of FILCT has been set at
43°C which is the target temperature in order to
reduce the tumor cell.
Figure 2 shows the micrograph of FILCT' and 'Au
coated FILCT’. FILCT was coated by Au in order to
enhance heat divergence efficiency. Au coated
FILCT was made using electroless gold plating
technique without underlaying harmful metal such
as Ni.
Figure 3 shows the block diagram of the
experimental setup of this experiment. Magnetic flux
was generated by the drive coil of 3 turns (External
diameter of the drive coil was 80mm, internal
diameter was 70mm). The frequency of the AC
current for the drive coil was 190 kHz. AC current
(500A) was generated by the power amplifier
(Hotshot, AMERITHERM Inc.). The shift of
vertical component of magnetic flux was measured
as variation of the induced voltage of the pickup coil.
The voltage of the pickup coil located orthogonally
in the drive coil was measured as the synchronous
signal of the voltage of the search coil by the lock-in
amplifier (7265 DSP Lock-in amplifier, Signal
Recovery). The number of turns, resistance, external
and internal diameter of the pickup coil was 10 turns,
10, 17mm and 10mm respectively. The
temperature of FILCT was measured by an optical
fiber thermometer (FL-2000, Anritsu-meter). In this
experiment, we used Au coated FILCT (particle
diameter was 40µm-150µm, total mass of powder
was 1.0g) with 500µl of deionized water. The
distance between the upper surface of the drive coil
and a test tube of FILCT was 30mm, and the density
of magnetic flux at the position of FILCT was
24.3mT.
Figure 1: The fundamental concept of less invasive
thermometry.
(a) FILCT (b) Au coated FILCT
Figure 2: Micrograph of a particle of the ferromagnetic
implant with low Curie temperature.
Power Unit
HOT SHOT
LOCK-IN AMPLIFIER
Pickup Coil
Seach Coil
FILCT
Drive Coil
Figure 3: Block diagram of the experimental setup of the
less-invasive heating and thermometry.
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
FILC T Sputter-deposited
FILC T
A u c o a t e d F IL C T
R a te o f tem perature increase [℃/s]
Figure 4: Rate of temperature increase.
BIODEVICES 2011 - International Conference on Biomedical Electronics and Devices
342
3 RESULTS AND DISCUSSIONS
Figure 4 shows the rate of temperature increase of
each implant. Applied density of magnetic flux was
7.85mT, the frequency was 190 kHz. The thickness
of ‘Au coated FILCT’ was around 0.7µm. The rate
of temperature increase of ‘Au coated FILCT’ was
decuple compared with original FILCT.
We repeated induction heating over 50 times in
order to evaluate the endurance of Au coating.
Figure 5 shows the temperature rise curve of ‘Au
coated FILCT’. Here, horizontal axis shows time
and vertical axis shows the average temperature of
50 times cycle test. As a result, the heat divergence
profile was constant. We did not find any change in
surface of Au coat FILCT such as a delamination or
a crack after cycle tests by microscopic observation.
Figure 6 shows the change of the temperature of
Au coated FILCT and the magnetic flux density at
the pickup coil. Vertical axis shows the temperature
of FILCT and the magnetic flux density at the
pickup coil. The self-heating type FILCT produced
heat of over 43 ˚C that was effective in the treatment
temperature of malignancy. The temperature of
FILCT rose to 43 ˚C within 30s. Rate of temperature
increase changed around the Curie temperature. The
magnetic flux density at the pickup coil was
decreased in contradiction to the rising of
temperature.
Figure 7 shows the relationship between the
temperature of FILCT and the voltage of pickup coil.
This graph was made from the data of figure 6 that
was the result of cyclic test of heating. Here,
horizontal axis shows the temperature and vertical
axis shows the voltage of pickup coil. The profile of
the voltage was almost same on each trial, and the
voltage was decreased to 6mV. At the same time, the
bias of voltage caused by the difference of position
of FILCT appeared.
This approach can pave the way to heating and
temperature measurements of tumor with no pain
after injecting of FILCT to the tumor. However,
long time measurements may be a burden to the
subject persons as the body movements are restricted
to minimize the noise for this method.
4 CONCLUSIONS
In this paper, we have described the concept for
using ‘ferromagnetic implant with low Curie
temperature (FILCT)’ as a probe for temperature. A
key advantage of this system is that it can
distinguish noninvasively whether the target
temperature has reached the Curie temperature or
not. Less invasive heating and temperature
measurement system was evaluated using powder
type of Au coated FILCT. As a result, it has become
clear that we can heat up to 50mm of depth distance
under thermal control. Furthermore, it is clear that
we can use Au coated FILCT which have high
heating efficiency for induction of heating. These
results contribute to the less invasive heating and
temperature measurement for a lot of types of
hyperthermia.
20
25
30
35
40
45
50
55
0 100 200 300 400 500 600
Time [s]
Temperature [
]
Figure 5: Temperature rise curve of 'Au coated FILCT'.
(50 times cycle tests).
25.0
30.0
35.0
40.0
45.0
50.0
55.0
0 20 40 60 80 100 120
Time [s]
Temperature [
]
119
120
121
122
123
124
125
126
127
128
129
130
Magnetic flux density a
t
Pickup Coil [μT]
Tem perature [℃]
M agnetic flux density at P ickup C oil [u T ]
Figure 6: Change of the temperature of Au-coated FILCT
and the voltage of pickup coil.
120
122
124
126
128
130
30.035.0 40.0 45.0 50.0
Tem perature [℃]
M agnetic flux density at
Pickup C oil T ]
Figure 7: Relationship between temperature of FILCT an
d
the voltage of the pickup coil.
LOW-INVASIVE HEATING AND TEMPERATURE MEASUREMENT METHOD FOR HYPERTHERMIA
TREATMENT USING THE METAL COATED FERROMAGNETIC IMPLANT WITH LOW CURIE TEMPERATURE
343
ACKNOWLEDGEMENTS
This work was supported in part by the Grant-in-Aid
for Scientific Research under Grant 21500438. We
express our thanks to Dr. Hajime Saito, an assistant
professor of medical school of Akita University. We
show our appreciation to Mr. Wu Yanfeng and Mr.
Yu Sugawara, graduated students of our laboratory.
REFERENCES
Abe M, Hiraoka M, Takahashi M et al., 1986. Multi-
institutional studies on hyperthermia using an 8-MHz
radiofrequency capacitive heating device (Thermotron
RF-8) in combination with radiation for cancer therapy.
Cancer, vol.58, pp.1589-95.
Cavaliere, R., Ciocatto, EC., Giovanella, BC., et al. 1967.
Selective heat sensitivity of cancer cells. Biochemical
and clinical studies. Cancer, vol.20, pp.1351-1381.
Hiraoka M, Nishimura Y, Nagata Y et al., 1994. Site-
specific phase I, II trials of hyperthermia at Kyoto
University. Int J Hyperthermia, Vol.10, pp.403-10.
Ishida, T., Kato, H., Miyakoshi, J., Furukawa, M., Ohsaki,
S. and Kano, E., 1980. Physical Basis of RF
Hyperthermia for Cancer Therapy (1) Measurement for
Distribution in Absorbed Power from Radiofrequency
Exposure in Agar Phantom, J. Radiat. Res., vol.21,
pp.180 -189.
Ito, A., Saito, H., Mitobe, K., Minamiya, Y., Takahashi,
N., Maruyama, K., Motoyama, S., Katayose, Y. and
Ogawa, J., 2009. Inhibition of heat shock protein 90
sensitizes melanoma cells to thermosensitive
ferromagnetic particle-mediated hyperthermia with
low curie temperature, Cancer Science, vol.100,
pp.558-64.
Jordan, A., Wust, P., Fahling, H., John, W., Hinz, A. and
Felix, R., 1993. Inductive heating of ferrimagnetic
particles and magnetic fluids: physical evaluation of
their potential for hyperthermia. Int J Hyperthermia,
vol.9, pp.51-68.
Matsuki, H., Murakami, K. and Niizuma, H., 1982. Soft
Heating - A New Method of Heating Using
Temperature-Sensitive Magnetic Cores, IEEE Trans.
Magn, vol.MAG-18, pp.1788-1790.
Mitobe, K. and Yoshimura, N., 2008. Noninvasive
Temperature Measurement Method for Hyperthermia
Treatment using Ferromagnetic Implant with Low
Curie Temperature, 30th Annual Inter. Conf. IEEE
EMBS, pp.4384-4386.
Overgaard, K. and Overgaard, J., 1972. Investigations on
the possibility of a thermic tumour therapy. I. Short-
wave treatment of a transplanted isologous mouse
mammary carcinoma. Eur J Cancer, Vol.8, pp.65-78.
Saito, H., Mitobe, K., Ito, A., Sugawara, Y., Maruyama,
K., Minamiya, Y., Motoyama, S., Yoshimura, N. and
Ogawa, J., 2008. Self-regulating hyperthermia induced
using thermosensitive ferromagnetic material having a
low Curie temperature, Cancer Science, No.99,
pp.805-809.
Shinkai, M., Yanase, M., Honda, H., Wakabayashi, T.,
Yoshida, J. and Kobayashi, T., 1996. Intracellular
hyperthermia for cancer using magnetite cationic
liposomes in vitro study. Jpn J Cancer Res, vol.87,
pp.1179-83.
BIODEVICES 2011 - International Conference on Biomedical Electronics and Devices
344