LOW NOISE MEASUREMENT OF PHOTOCURRENT FOR CONTINUOUS GLUCOSE MONITORING - Low Noise Measurement System Enables Continuous Monitoring of Glucose in Subcutaneous Interstitial Fluid

Daniel W. Cooley, David R. Andersen

2010

Abstract

We have developed a data acquisition unit (DAU) for continuous, low noise measurement of glucose concentration in subcutaneous interstitial fluid (ISF). The system is comprised of a glucose sensor (Olesberg, 2006), op-amps for signal conditioning, and delta-sigma (∆-Σ) ADCs. The glucose sensor has two IR LEDs which emit light with wavelength of 2.2 to 2.4 μm where there are peaks in the glucose absorption spectrum. The IR light propagates through a glass fluid chamber containing interstitial fluid and a linearly variable bandpass filter before impinging on a 32 channel photodiode array. The center frequency of the filter varies along one dimension of the filter, so that each photodiode is sensitive to equally spaced portions of the glucose absorption spectrum. Transimpedance amplifiers (TIAs) convert the photocurrents into voltages which are sampled by ADCs. We developed a noise model which predicts the noise characteristics of the system. We use low noise metal film resistors to verify the DAU noise characteristics. Non-ideal characteristics such as limited photocurrent and low photodiode shunt resistance increase difficulty of obtaining low noise measurements. We demonstrate that the DAU provides low noise (41.7 dB SNR) photocurrent measurements and is suitable for use in a continuous glucose monitoring system.

References

  1. Olesburg, J. T., Cao, C., Yager, J. R., Prineas, J. P., Coretsopoulos, C., Arnold, M. A., L. J. Olafsen, L. J., Santilli, M., 2006. Proc. of SPIE 694, 609403.
  2. Centers for Disease Control and Prevention, 2007. "National Diabetes Fact Sheet, 2007,” Department of Health and Human Services - Centers for Disease Control and Prevention.
  3. Saudek, C. D., Derr, R. L., Kalyani, R. R., 2006. JAMA 295 No. 14 p. 1688.
  4. Chia, C. W., Saudek, C. D., 2004. Endocrinol. Metab. Clin. N. Am. 33 p. 175.
  5. Rebrin, K., Steil, G. M., 2000. Diabetes Tech. and Therapeutics 2 No. 3 p. 461.
  6. Motchenbacher, C. D., Connelly, J. A., 1993. Low-Noise Electronic System Design, John Wiley and Sons. New York.
  7. Steffes, M., 2005. Application Report SBOA066A, "Noise Analysis for High-Speed Op Amps,” (www.ti.com).
  8. Hamamatsu, 2001. Datasheet, “Si Photodiode S1087/S1133 series,” Hamamatsu Photonics K. K.
  9. Maxim, 2005. MAX4478 datasheet. “MAX4478 SOT23, Low-Noise, Low-Distortion, Wide-Band, Rail-to-Rail Op Amps,” Maxim Integrated Products, Inc.
  10. Texas Instruments, 2008. ADS1278 datasheet. “ADS1278 Quad/Octal, Simultaneous Sampling, 24-Bit Analogto-Digital Converters,” Texas Instruments, Inc.
  11. Cirrus Logic, 2009. VRE3025 datasheet. “VRE3025 Precision Voltage Reference,” Cirrus Logic, Inc.
Download


Paper Citation


in Harvard Style

W. Cooley D. and R. Andersen D. (2010). LOW NOISE MEASUREMENT OF PHOTOCURRENT FOR CONTINUOUS GLUCOSE MONITORING - Low Noise Measurement System Enables Continuous Monitoring of Glucose in Subcutaneous Interstitial Fluid . In Proceedings of the Third International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2010) ISBN 978-989-674-017-7, pages 195-200. DOI: 10.5220/0002747901950200


in Bibtex Style

@conference{biodevices10,
author={Daniel W. Cooley and David R. Andersen},
title={LOW NOISE MEASUREMENT OF PHOTOCURRENT FOR CONTINUOUS GLUCOSE MONITORING - Low Noise Measurement System Enables Continuous Monitoring of Glucose in Subcutaneous Interstitial Fluid},
booktitle={Proceedings of the Third International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2010)},
year={2010},
pages={195-200},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0002747901950200},
isbn={978-989-674-017-7},
}


in EndNote Style

TY - CONF
JO - Proceedings of the Third International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2010)
TI - LOW NOISE MEASUREMENT OF PHOTOCURRENT FOR CONTINUOUS GLUCOSE MONITORING - Low Noise Measurement System Enables Continuous Monitoring of Glucose in Subcutaneous Interstitial Fluid
SN - 978-989-674-017-7
AU - W. Cooley D.
AU - R. Andersen D.
PY - 2010
SP - 195
EP - 200
DO - 10.5220/0002747901950200