loading
Papers Papers/2022 Papers Papers/2022

Research.Publish.Connect.

Paper

Paper Unlock

Authors: Markus Valtin 1 ; Thomas Schauer 2 ; Carsten Behling 3 ; Michael Daniel 3 and Matthias Weber 3

Affiliations: 1 Technische Universität Berlin and HASOMED GmbH, Germany ; 2 Technische Universität Berlin, Germany ; 3 HASOMED GmbH, Germany

Keyword(s): Multi-channel transcutaneous electrical stimulation, Array electrode, Electromyography.

Related Ontology Subjects/Areas/Topics: Biomedical Engineering ; Biomedical Equipment ; Biomedical Instrumentation ; Biomedical Instruments and Devices ; Emerging Technologies ; Telecommunications ; Wireless and Mobile Technologies ; Wireless Information Networks and Systems

Abstract: Array electrodes have the potential to significantly advance Functional Electrical Stimulation (FES) performance and patient compliance by optimizing the electrode position. To evaluate the potentials and for research purposes, an universal stimulation system for array electrodes has been developed. The system additionally features volitional EMG recording from the array electrodes during active stimulation. Multiple devices, one stimulator and at least one demultiplexer, are synchronized to deliver up to 10 stimulation pulses per stimulation cycle at a frequency of  420Hz. A typical stimulation cycle period is 50ms. The real-time controllable array electrode can include up to 60 elements for the active electrode and up to 4 elements for the indifferent electrode. A small switch module permits placement near the array electrode, eliminating extensive wiring. The stimulation system is fully controllable from a PC via USB interfaces.

CC BY-NC-ND 4.0

Sign In Guest: Register as new SciTePress user now for free.

Sign In SciTePress user: please login.

PDF ImageMy Papers

You are not signed in, therefore limits apply to your IP address 216.73.216.157

In the current month:
Recent papers: 100 available of 100 total
2+ years older papers: 200 available of 200 total

Paper citation in several formats:
Valtin, M., Schauer, T., Behling, C., Daniel, M., Weber and M. (2012). COMBINED STIMULATION AND MEASUREMENT SYSTEM FOR ARRAY ELECTRODES. In Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2012) - BIODEVICES; ISBN 978-989-8425-91-1; ISSN 2184-4305, SciTePress, pages 345-349. DOI: 10.5220/0003786303450349

@conference{biodevices12,
author={Markus Valtin and Thomas Schauer and Carsten Behling and Michael Daniel and Matthias Weber},
title={COMBINED STIMULATION AND MEASUREMENT SYSTEM FOR ARRAY ELECTRODES},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2012) - BIODEVICES},
year={2012},
pages={345-349},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0003786303450349},
isbn={978-989-8425-91-1},
issn={2184-4305},
}

TY - CONF

JO - Proceedings of the International Conference on Biomedical Electronics and Devices (BIOSTEC 2012) - BIODEVICES
TI - COMBINED STIMULATION AND MEASUREMENT SYSTEM FOR ARRAY ELECTRODES
SN - 978-989-8425-91-1
IS - 2184-4305
AU - Valtin, M.
AU - Schauer, T.
AU - Behling, C.
AU - Daniel, M.
AU - Weber, M.
PY - 2012
SP - 345
EP - 349
DO - 10.5220/0003786303450349
PB - SciTePress