ADIPSMETER - A New Skinfold Calliper System

Maria de Fátima Chouzal, Maria Teresa Restivo, Manuel Rodrigues Quintas, Carlos Moreira da Silva, Tiago Andrade, Maria Teresa Amaral

2011

Abstract

Nowadays, the assessment of body composition is of great meaning but traditional skinfold callipers recognized in the nutritional area have not been technologically updated to provide a more effective performance with greater accuracy, allowing automatic data processing and recording in a database, offering a more user friendly handling not requiring technicians with a high degree of expertise. In order to overcome many of these deficiencies a new instrument named Adipsmeter was designed and developed. A new mechanical skinfold calliper structure has been digitally instrumented for reading skinfold thickness and equipped with wireless communication capabilities with a personal computer where a software application, integrating a database, is installed and prepared for driving a technician along a specific protocol incorporating many regression equations used in current body fat measurement at an affordable price. Its measurement range is from 0 to 110 mm, consistent with a range of use from children to obese individuals. Between the two opposing clamping surfaces mounted on its rigid arms the pressure has a constant value of 10 gf/mm2 in the whole measurement range. It has small dimension and weight and while being robust provides comfortable handling. The sensing system resolution is less than 0.1 mm, the wireless communication (MiWi) is suitable for healthcare environments, with 25 h of autonomy. The software application, with an intuitive and user friendly interface, is based in Visual Basic 2008.

References

  1. Amaral, T., Restivo, M. T., Guerra, R., Marques, E., Chousal, M. F., Mota, J. (2010). Validation of a digital skinfold system for estimating body fat based on skinfold thickness measurement. Accepted to the British Journal of Nutrition.
  2. Barbosa, M. R., Amaral, T., Chouzal, M. F., Restivo, M. T. (2010). Neural Networks Based Approach to Estimate Body Fat (%BF). In Controlo2010, 7-10 September, Coimbra Portugal.
  3. Durnin J. V., Womersley J. (1974). Body fat assessed from total body density and its estimation from skinfold thickness: measurements on 481 men and women aged from 16 to 72 years. British Journal of Nutrition; 32(1):77-97.
  4. European Parliament. (2008). European Parliament Resolution of 25 September 2008 on the White Paper on nutrition, overweight and obesity-related health issues (2007/2285(INI)) P6_TA(2008)0461.
  5. Jackson A. S., Ellis K. J., McFarlin B. K. et al. (2009). Cross-validation of generalized body composition equations with diverse young men and women: the Training Intervention and Genetics of Exercise Response (TIGER) Study. British Journal of Nutrition 101, 871-8.
  6. Lohman T. G. (1992). Advances in Body Composition Assessment: Current Issues in Exercise Science, Champaign, IL: Human Kinetics: 3-4. Series Monograph No. 3.
  7. Lohman T. G. (1996). Dual energy X-ray absorptiometry in human body composition. eds. A Roche, S Heymsfield & T G Lohman, pp 63-78. Champaign, IL: Human Kinetics.
  8. Norton K, Olds T. (1996). Measuring Techniques in anthropometry - anthropometry equipment. In Anthropometrica, pp. 29-32. Sydney, Australia: University of New South Wales Press.
  9. Siri W. E. (1961). Body composition from fluid spaces and density. Analysis of methods, In Techniques for Measuring Body Composition, Edited by: Brozek J, Henschel A. Washington, DC , National Academy of Sciences.
  10. Restivo, M. T., Amaral, T., Mendes, J. G., Quintas, M. R. and Chouzal, M. F. (2009). Dispositivo para aquisição e processamento de dados para determinação da massa corporal. Boletim da Propriedade Industrial nº 2009/11/26.
  11. Restivo, M. T., Amaral, T., Chousal, M. F., Leão, C. P., Guerra, R., Marques, E., Mendes, J., Quintas, M. and Mota, J. (2010). Skinfold calliper integrating new technologies. Submitted to the Measurement Science & Technology Journal.
Download


Paper Citation


in Harvard Style

de Fátima Chouzal M., Teresa Restivo M., Rodrigues Quintas M., Moreira da Silva C., Andrade T. and Teresa Amaral M. (2011). ADIPSMETER - A New Skinfold Calliper System . In Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2011) ISBN 978-989-8425-37-9, pages 174-178. DOI: 10.5220/0003140101740178


in Bibtex Style

@conference{biodevices11,
author={Maria de Fátima Chouzal and Maria Teresa Restivo and Manuel Rodrigues Quintas and Carlos Moreira da Silva and Tiago Andrade and Maria Teresa Amaral},
title={ADIPSMETER - A New Skinfold Calliper System},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2011)},
year={2011},
pages={174-178},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0003140101740178},
isbn={978-989-8425-37-9},
}


in EndNote Style

TY - CONF
JO - Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2011)
TI - ADIPSMETER - A New Skinfold Calliper System
SN - 978-989-8425-37-9
AU - de Fátima Chouzal M.
AU - Teresa Restivo M.
AU - Rodrigues Quintas M.
AU - Moreira da Silva C.
AU - Andrade T.
AU - Teresa Amaral M.
PY - 2011
SP - 174
EP - 178
DO - 10.5220/0003140101740178