MRI-INDUCED SAR ON PACEMAKER LEADS - Numerical Simulations on Three Human Phantoms

Eugenio Mattei, Giovanni Calcagnini, Michele Triventi, Federica censi, Pietro Bartolini

2011

Abstract

Numerical simulations were performed to evaluate the Specific Absorption Rate (SAR) induced at the tip of a pacemaker (PM) implant by the 64 MHz radiofrequency (RF) field used in 1.5T Magnetic Resonance Imaging (MRI) procedures. The analysis was performed by using a commercial FDTD software (SEMCAD X, SPEAG, Switzerland) and aimed at the evaluation of the impact that the patient ‘s morphology has on the induced local SAR at the implant tip. In particular three human phantoms were studied: a 34-year old man model, a 26-year old woman, and a 6-year old boy. The three phantoms reproduce more than 70 tissues of the human body with a resolution of 1 mm. Inside each phantoms, realistic implant configurations were modelled, considering both left and right pectoral implants, and atrial and ventricular stimulations. The local SAR values at the lead tip was compared for the three phantoms and sensible differences were observed: with a RF excitation set to produce an whole-body average SAR of 2 W/kg without any implants, local SAR values ranged from 641W/kg (woman model – right ventricular implant) to 3 W/kg (boy model – left atrium implant). We also observed that, in general, ventricular implants showed a higher SAR compared to atrial ones, as well as right pectoral implants compared to left ones. However, not always a higher implant area or a longer lead path implied higher SAR at the tip, indicating the coupling mechanisms between the implant and the RF field are likely to be more complex that the only area-dependent induction law.

References

  1. ASTM F2182-02a - "Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging", ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA, 19428-2959 USA
  2. IEEE, IEEE1528, recommended practice for determining the spatial-peak specific absorption rate (SAR) in the human body due to wireless communications devices: Measurement techniques, IEEE Standards Department 52(5), 1 2006
  3. Mattei E, Triventi M, Calcagnini G, Censi F, Kainz W, Mendoza G, Bassen HI, Bartolini P. Complexity of MRI induced heating on metallic leads: experimental measurements of 374 configurations Biomed Eng Online. 2008 Mar 3;7:11
  4. Nordbeck P, Weiss I, Ehses P, Ritter O, Warmuth M, Fidler F, Herold V, Jakob PM, Ladd ME, Quick HH, Bauer WR. Measuring RF-induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leads. Magn Reson Med. 2009 Mar;61(3):570-8.
  5. Rezai AR, Baker KB, Tkach JA, Phillips M, Hrdlicka G, Sharan AD, Nyenhuis J, Ruggieri P, Shellock FG, Henderson J. Is magnetic resonance imaging safe for patients with neurostimulation systems used for deep brain stimulation? Neurosurgery 2005 Nov;57(5):1056-62
  6. Roguin A, Zviman MM, Meininger GR, Rodrigues ER, Dickfeld TM, Bluemke DA, Lardo A, Berger RD, Calkins H, Halperin HR. Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe: in vitro and in vivo assessment of safety and function at 1.5 T. Circulation. 2004 Aug 3;110(5):475-82.
  7. Sommer T, Vahlhaus C, Lauck G, von Smekal A, Reinke M, Hofer U, Block W, Traber F, Schneider C, Gieseke J, Jung W, Schild H. MR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 T. Radiology 2000 Jun;215(3):869- 79
Download


Paper Citation


in Harvard Style

Mattei E., Calcagnini G., Triventi M., censi F. and Bartolini P. (2011). MRI-INDUCED SAR ON PACEMAKER LEADS - Numerical Simulations on Three Human Phantoms . In Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2011) ISBN 978-989-8425-37-9, pages 135-139. DOI: 10.5220/0003131501350139


in Bibtex Style

@conference{biodevices11,
author={Eugenio Mattei and Giovanni Calcagnini and Michele Triventi and Federica censi and Pietro Bartolini},
title={MRI-INDUCED SAR ON PACEMAKER LEADS - Numerical Simulations on Three Human Phantoms},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2011)},
year={2011},
pages={135-139},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0003131501350139},
isbn={978-989-8425-37-9},
}


in EndNote Style

TY - CONF
JO - Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2011)
TI - MRI-INDUCED SAR ON PACEMAKER LEADS - Numerical Simulations on Three Human Phantoms
SN - 978-989-8425-37-9
AU - Mattei E.
AU - Calcagnini G.
AU - Triventi M.
AU - censi F.
AU - Bartolini P.
PY - 2011
SP - 135
EP - 139
DO - 10.5220/0003131501350139