Principles of Organization and Control of the New Implementation of
the “SYMPATHOCOR-01” Neuro-electrostimulation Device
Vladimir Kublanov, Mikhail Babich and Anton Dolganov
Research Medical and Biological Engineering Centre of High Technologies, Ural Federal University, Mira 19, 620002,
Yekaterinburg, Russian Federation
Keywords: Neuro-electrostimulation, Mobile Device, Device Control, Device Organization, “SYMPATHOCOR-01”
Abstract: The paper describes peculiarities of the organizational and control principles of the new implementation of
the “SYMPATHOCOR-01” device. Necessity of the device division into two blocks, having different
functions, was justified. The specifics of the schematic implementation were mentioned. Advantages of the
wireless connection between two blocks were shown. Particular possibilities and actual implementation of
the control block via Android application for mobile devices were presented. The potential targets for the
neuro-electrostimulation that affect autonomic processes were considered.
1 INTRODUCTION
It is well known that combination of several kinds of
stimulation, each of which provides different types
of physiological reactions of damaged cognitive,
autonomic, motor, sensory and vestibular functions,
achieves the best efficiency in the process of
neurorehabilitation. This is necessary for the specific
attraction of new neuronal resources into the
damaged neural network, which provides
rehabilitated function (Enriquez-Geppert et al.,
2013). This principle has been used for a long time
in the creation of the effective programs for the
physical rehabilitation by means of the
transcutaneous electrical muscle stimulation (Nagai
et al., 2016).
According to modern science concepts, addition
of the physical and cognitive load to the process of
neurorehabilitation can increase the production of
neurotrophic factors and activate the mirror neuron
system (Yuan et al., 2015). The neurotrophic factors
implement the neuroplasticity of the brain
mechanisms. At the same time, mirror neurons are
able to remodel lost functions (Reichardt, 2006).
Devices that are compact, portable and do not
cause discomfort for patients during rehabilitation
procedures are the most promising for the
combination of several kinds of stimulation during
treatment process. At present time, many neuro-
electrostimulation devices that meet these
requirements are designed.
Neuro-electrostimulation devices, which use
low-frequency current pulses with single polarity,
are among the perspectives types of the
neuroelectrostimulation. The duration and frequency
of such devices commensurate with the duration of
nerve pulses and the frequency of their appearance
in neural networks.
The number of corresponding devices decreases
dramatically, if one exclude devices with implanted
electrodes. Non-invasive neuro-electrostimulation
devices, that include multi-electrode systems, have
greater opportunities for activation of the brain
neuroplasticity mechanisms, compared to other
devices, due to the redundancy of the information
generated by their spatially distributed current pulses
fields.
Two original multi-channel portable systems for
non-invasive stimulation of the tongue (PoNS
device) (Danilov and Kublanov, 2014) and the neck
area (“SYMPATHOCOR-01” device) (Kublanov
2008) are fully satisfied of these requirements.
The “SYMPATHOCOR-01” device is included
in the register of medical equipment products of the
Russian Federation and has the Certificate of
correspondence to the requirements of the Safety.
The Manual for its use was approved by the State
Control of the quality, effectiveness and safety of the
medicines and medical equipment Department of the
Russian Ministry of Health.
This article discusses the features new technical
Kublanov V., Babich M. and Dolganov A.
Principles of Organization and Control of the New Implementation of the â
˘
AIJSYMPATHOCOR-01â
˘
A
˙
I Neuro-electrostimulation Device.
DOI: 10.5220/0006175802760282
Copyright
c
2017 by SCITEPRESS – Science and Technology Publications, Lda. All rights reserved
implementation of the compact and mobile
“SYMPATHOCOR-01” device. The device is aimed
for application in neurorehabilitation tasks that
imply using of several kinds of stimulation.
2 NEURO-
ELECTROSTIMULATION
TARGETS
The basis of all human mental activity are the
processes in the central nervous system. It should be
noted that the role of the cerebral circulation: mental
performance (attention, memory and perception,
logical thinking) is reduced at the deterioration of
blood supply to the brain. This feature determines
the search for solutions to manage the blood supply
of the brain. Therefore, those physiological
mechanisms of the sympathetic nervous system are
fundamental which allows to control the tone of the
blood vessels of different caliber.
The most important formations that are involved
in the organization of the neuro-electrostimulation
are:
glossopharyngeal nerve and its branches;
vagus nerve and its branches;
the accessory nerve;
the nerve plexus around the carotid artery;
the sympathetic trunk structures
o upper cervical node;
o middle cervical node;
o vertebral ganglion;
o stellate ganglion;
spinal nerves C2-C4 forming the cervical
plexus.
Figure 1 shows the conventional areas of the
nerve structures location in the neck region.
Regulating centers of the vital functions are
placed in the nuclei of the brain stem, midbrain,
pons and the cerebellum, in the autonomic nuclei of
the brain and spinal cord. Many of the mentioned
pathways are located in the neck.
The nervous formations of neck area are closely
associated with brainstem, which have two-side
connections with midbrain, cerebellum, thalamus,
hypothalamus and the large brain cortex. Presence of
these connections provides participation of the neck
nervous formations in analysis of sensory
stimulation, regulation of the muscle tonus,
autonomic and the highest integrative functions
(Netter, 2010).
Figure 1: Conventional areas of the nerve structures
location in the neck: 1 – n. glossopharyngeus; 4 – n.
plexus caroticus externus; 11 – ganglion cervicale
medium; 15 – ganglion stellatum; 21 – ramus
communicans griseus; 22 – ganglion cervicale superius;
23 – n. vagus.
As a stimulation targets can be used:
the superior cervical ganglia of the
sympathetic nervous system;
the stellate ganglion;
other components of the sympathetic trunk;
the afferent branches of the cervical plexus;
cranial nerves and their branches (IX, X and
XI pair).
These targets are the conductive paths of the
brain stem nerve structures. Usage of them
significantly extends the capabilities of the neuro-
electrostimulation method (Kublanov and Babich,
2015).
The stimulation of neck nodes of the sympathetic
trunk affects both the vascular tone of arteries of the
brain, and autonomic spinal nucleus. Thus, neuro-
electrostimulation device is able to fully modulate
the autonomic processes and affect motor control
and cognitive functions.
3 ORGANIZATION OF THE
“SYMPATHOCOR-01” DEVICE
The “SYMPATHOCOR-01” is the non-invasive
neuro-electrostimulation device that performs
stimulation by the spatially distributed current pulses
field using two multi-electrode (ME) systems. ME
systems are placed symmetrically on the neck skin,
in projections of the sympathetic nervous system
neck ganglion.
Partial current pulses that flows from partial
cathodes of one ME system to single anode of the
other ME system carry out the stimulation process.
Partial current pulses have the amplitude, ranging
from 0 to 15 mA; the duration, ranging from 15 to
60 us; the modulation frequency from 5 to 150 Hz.
For the neuro-electrostimulation process from 1 to
13 partial cathodes can be involved.
3.1 General Structure
The “SYMPATHOCOR-01” device consists of two
blocks. First block has autonomous power source
(built-in Li-ion battery) and forms the spatially
distributed field of current pulses. The energy
consumption of the first block reaches 47 mA. Time
of the continuous operation reaches 22 hours.
Second block sets characteristics of the spatially
distributed current pulses field, field’s configuration
in the neck area, based on functional state of the
autonomic and central nervous systems of the
patient. Personal computer or mobile device can be
used as the second block.
Fig. 2 presents the block-scheme of the
“SYMPATHOCOR-01” device.
Figure 2: Block-scheme of the “SYMPATHOCOR-01”
device.
The built-in processors control both blocks. The
wireless communication channel provides
connection between two blocks. Each block has a
built-in receiver-transmitter. The Bluetooth Low
Energy technology is used as the wireless
communication channel. The second block
implement virtual block of the characteristics setting
as the graphical user interface accessible to the
doctors.
The wireless communication channel between
two blocks has possibility to perform the
information transfer using unique addresses of the
receiver-transmitters, placed in both blocks.
Therefore, it is possible for one doctor to organize
management of simultaneous neuro-
electrostimulation for group of patients by means of
single second block.
The block-scheme of the
“SYMPATHOCOR-01” organization in case of
simultaneous neuro-electrostimulation of two
patients is shown on Fig. 3.
Figure 3: Block-scheme of the “SYMPATHOCOR-01”
device organisation for two patients.
3.2 The First Block Structure
The division of the “SYMPATHOCOR-01” device
into two blocks, in comparison with older releases,
allowed to exclude user interface used by doctor,
and to simplify the processor installed in the first
block. Simplification of the first block processor
arisen from shifting of particular functions to the
second block. Because of that the first block was
implemented as mobile and compact device with the
following dimensions: 90.0 50.0 18.5 mm. The
general view of the “SYMPATHOCOR-01” device
first block is shown on Fig. 4. Fig. 5 shows general
view of the main printed circuit board (PCB) in the
“SYMPATHOCOR-01” device first block.
8-channel chips of the Analog Devices
ADG5408 multiplexers are the basis for
commutation of the partial cathodes and anodes. In
the present implementation of the neuro-
electrostimulation device, four chips are used. In this
case, it is possible to use in the stimulation process
up to 16 partial cathodes and 16 anodes.
Three bipolar transistors (two BC807 and one
BC817) implement current source of the neuro-
electrostimulation by means of the current mirror.
The built-in 12-bit digital-to-analog converter
generates control voltage of the current source. The
resistor values of the current mirror were selected in
a way that current source generates current with the
amplitude in a range from 0 to 15 mA, with load in
range from 0 to 2 kOhm.
The algorithm of the current pulses field
formation, implemented in the first block processor
is shown on Fig. 6.
Figure 4: General view of the first block.
Figure 5: General view of the first block printed circuit
board.
Figure 6: The algorithm of the field formation.
Change of the current pulses field state is
possible only in certain discrete time points. These
time points are defined by the equation
=
+∗,
(1)
where: ∈, ∈,0, K – the number of
the partial cathodes used in the neuro-
electrostimulation, – the duration of the partial
pulse, ν – the modulation frequency of the current
pulse field.
The processor performs the following steps for
i-th time point:
Consequent turning off the previously
connected cathode and turning on the
following cathode in accordance with the
neuro-electrostimulation program;
Changing of the anodes, in case of the
respective requirements of the neuro-
electrostimulation program or by doctor
command;
In case of new target values of the amplitude,
the duration, the modulation frequency of the
partial current pulses, set by the second block
of the device, changing of the respective
characteristics in accordance with the
following equations:
i
=
targe
t
,
i−1
targe
t
<∆
A
i−1
+∆
A
,
i−1
+∆
A
<
target
i−1
−∆
A
,
i−1
−∆
A
>
targe
t
(2)
i
=
targe
t
,
i−1
−
targe
t
<∆
T
i−1
+∆
T
,
i−1
+∆
T
<
target
i−1
−∆
T
,
i−1
−∆
T
>
target
=1/
(3)
τ
i
=
τ
targe
t
,τ
i−1
−τ
targe
t
<∆
τ
τ
i−1
+∆
τ
i−1
+∆
τ
target
τ
i−1
−∆
τ
i−1
−∆
τ
targe
t
,
(4)
where: A – the amplitude of the partial current
pulses, A
target
– the target amplitude of the partial
current pulses, Δ
A
– the maximum amplitude
increment, T
target
– the target period of the current
pulses, ΔT – the maximum period increment,
τ
target
– the target duration of the partial pulse, Δ
τ
the maximum duration increment.
Usage of the constraint to the rate of the
characteristics change allowed preventing pain
feelings during the neuro-electrostimuation process.
3.3 The Second Block Structure
Currently, the second block control program is
implemented as the application for the mobile
devices with the operational system Android 4.2.2
and higher. Fig. 7 shows general view of the
graphical user interface for the application.
The application allows one to control and change
the amplitude, duration and the modulation
frequency of the partial current pulses, to switch the
currently used anode and to change field direction in
real time.
The characteristics of the partial current pulses
can be set by either using the seek bar or through
direct value setting in the corresponding fields, to
the right of seek bar elements. For the amplitude
changes buttons of coarse (‘+1’, ‘-1’) and fine
(‘+0.1’, ‘-0.1’) change were additionally
implemented. This addition allowed preventing pain
feelings caused by unreasonable amplitude increase.
Figure 7: General view of the graphical user interface.
The group of radio buttons are used for switching
of the current anode. The pattern of the radio buttons
placement is similar to the pattern of the partial
electrodes placement of the ME systems.
The filed state can be either ‘Turned off’, or
‘Direct’ or ‘Reverse’. Three corresponding buttons
on the graphical user interface control the field state.
The ‘Reverse’ state changes role of the ME systems.
In present time, two modes of the currently used
anode switch are implemented: ‘Manual’ and
‘Automatic’. In case of manual mode, the doctor
switch currently used anode based on functional
state of the patient. In case of automatic mode,
currently used anode switch in accordance with the
specific neuro-electrostimulation algorithm.
4 RESULTS
In the present time, the efficacy of the
“SYMPATHOCOR-01” device was proven both, for
prenosological state of the pathology development
and for patients treatment, suffering from the
disorders of the cognitive, autonomic, motor, sensor
and vestibular functions. The
“SYMPATHOCOR-01” device is used in more than
200 Russian hospitals (Kublanov et al., 2010). The
device is applied for correction of the following
pathologies:
migraine;
neurocirculatory dystonia;
alcohol and narcotic abstinences;
hypertonic diseases;
obliterate atherosclerosis of the lower limbs;
Raynaud’s disease;
trigeminal nerve inflammation;
sensorineural deafness;
degenerative diseases of vision;
atrophy of the optical nerve;
osteochondrosis of the back bone;
neuropathies of various genesis;
cephalalgia syndrome;
hyperhidrosis syndrome;
syndrome of the orthostatic hyposthenia;
postural tachycardia;
vestibular disease;
vegetative deregulation syndrome;
epilepsy;
attention deficit hyperactivity disorder.
New mobile implementation of the
“SYMPATHOCOR-01” device can be used for the
neuro-electrostimulation in combination with
different kinds of the stimulation. This extends the
application of the device for the following
directions:
treatment of the cognitive disorders caused by
the cranial injuries, strokes, neuroinfections
during the athrophic diseases of the human
brain;
treatment of the depressive and anxiety
disorders;
improvement of the senso-motor coordination
and recovery of the autonomic functions for
patients with brain circulation disorders;
rehabilitation of the military personnel;
target transport of the drugs in the local areas
of the organism;
preventative medicine;
recovery of the sportsmen in the training
period and during the competitions in the sport
medicine.
Table 1 presents comparison of electrical and
ergonomic characteristics of the new and old
technical implementations of the
“SYMPATHOCOR-01” device. The standard
laboratory equipment, which includes oscilloscopes
and the power supply, were used for comparison of
the characteristics. A system of resistors, that
connects all cathodes and all anodes at the same
point, was used as an equivalent to biological
material.
Table 1: Comparison of characteristics of the new and old
technical implementations of the “SYMPATHOCOR-01”
device.
Characteristic
Old technical
implementation
New technical
implementation
Number of
partial cathodes
12 13
Number of
partial anodes
1 13
Weight 1.5 kg 200 g
Dimensions,
mm
186x116x60 90x50x18,5
Amplitude of
the partial
pulses
0 - 25 V 0 - 15 mA
Frequency of
the pulse field
modulation, Hz
15 – 100 5 – 150
Length of the
partial pulse, µs
35 – 50 15 – 60
Set of the field’s
parametres
Switchers on the
panel of the
device
Wireless
channel by
means of
Bluetooth Low
Energy
Also device with new technical implementation
is capable of changing program of the neuro-
electrostimulation and conducting procedures on
multiple patients by single second block.
Comparative analysis of the data in table 1 shows
the increase of the possibilities of the device in
neuroreabilitation tasks due to decrease of weight
and device proportions. New implementation of the
device does not cause any discomfort to patient
during the procedures. In addition to that,
application of the wireless channel simplify
combination of the neuro-electrostimulation with
other kinds of stimulation used in the rehabilitation.
The ability to change program of the neuro-
electrostimulation as well as availability of the
multiple partial anodes allows to involve in the
stimulation process not only neck ganglia, but also
influence on the stellate ganglion, the afferent
branches of the cervical plexus, cranial nerves and
their branches as well as other components of the
sympathetic trunk. All in all, abovementioned
improvements significantly enhance possibilities of
the new technical implementations of the
“SYMPATHOCOR-01” device in
neurorehabilitation tasks.
5 CONCLUSIONS
The article describes new technical implementation
of the “SYMPATHOCOR-01” device.
Implementation of the device as the two block
allowed to improve ergonomic characteristics, to
make the device mobile and compact, to realize the
simultaneous stimulation of the patients group by
single doctor. Presented in the article analysis of the
organizational and control principles of the
“SYMPATHOCOR-01” device new implementation
revealed the innovative possibilities in the
neurorehabilitation tasks. Device mobility and
wireless control allow doctors to combine several
kind of stimulation including cognitive and motor
loads and neuro-electrostimulation using
“SYMPATHOCOR-01” device.
The next step of device implementation is
automatic real-time recording of the functional
changes in the central and autonomic nervous
systems. This recording will allow us to improve
informational control of the treatment management
and to increase treatment efficiency.
ACKNOWLEDGMENTS
The work was supported by Act 211 Government of
the Russian Federation, contract 02.A03.21.0006.
REFERENCES
Danilov, Y. P. and Kublanov, V. S., 2014. Emerging
Noninvasive Neurostimulation Technologies: CN-
NINM and SYMPATOCORECTION. Journal of
Behavioral and Brain Science, 4 (3), 105–113.
Enriquez-Geppert, S., Huster, R.J., and Herrmann, C.S.,
2013. Boosting brain functions: Improving executive
functions with behavioral training, neurostimulation,
and neurofeedback. International Journal of
Psychophysiology, 88 (1), 1–16.
Kublanov, V. S., 2008. A hardware-software system for
diagnosis and correction of autonomic dysfunctions.
Biomedical Engineering, 42 (4), 206–212.
Kublanov, V. S. and Babich, M. V., 2015. Principles of
organization and control of multielectrode neuro-
electrostimulation device. In: Biomedical Engineering
and Computational Technologies (SIBIRCON), 2015
International Conference on. IEEE, 82–86.
Kublanov, V. S., Shmirev, V. I., Shershever, A. S., and
Kazakov, J.E., 2010. About Innovative Possibilities of
Device ‘SIMPATOCOR-01’ in Management of
Functional Disorders of Vegetative and Central
Nervous System in Neurology. Kremlin medicine.
Clinical Vestnik, 4, 60–64.
Nagai, M. K., Marquez-Chin, C., and Popovic, M. R.,
2016. Why Is Functional Electrical Stimulation
Therapy Capable of Restoring Motor Function
Following Severe Injury to the Central Nervous
System? In: M. H. Tuszynski, ed. Translational
Neuroscience. Springer US, 479–498.
Netter, F. H., 2010. Atlas of human anatomy. Elsevier
Health Sciences.
Reichardt, L.F., 2006. Neurotrophin-regulated signalling
pathways. Philosophical Transactions of the Royal
Society of London B: Biological Sciences, 361 (1473),
1545–1564.
Yuan, T.-F., Chen, W., Shan, C., Rocha, N., Arias-
Carrion, O., Paes, F., de Sa, A.S., and Machado, S.,
2015. Activity-Dependent Neurorehabilitation Beyond
Physical Trainings: ‘Mental Exercise’ Through Mirror
Neuron Activation. CNS & Neurological Disorders -
Drug Targets (Formerly Current Drug Targets, 14
(10), 1267–1271.